Simple Questions:
| The machine may have quit
pumping but still getting heat, check to see if fluid is warm. If
yes, go to no flow.
Proceed if no heat nor pump. |
|
| Check to see if the outlet is good
(simply by plugging something else into it) |
|
| If there is a GFCI switch on the power cord,
is it popped? If so, reset switch. |
|
|
- Still no power, proceed to machine type:
|
|
.Replace Control Panel.
It's possible (however unlikely) that it could also be
the power cord. If you'd like to be sure, drop the panel,
and measure voltage coming in from power cord. Panel can be
disconnectd by unscrewing 4 screws in corner of control panel.
- If you get a power reading, control panel is bad.
- If no reading (but know the outlet is good) the
power cord is bad.
|
Back to top
.It's possible (however
unlikely) that it could be the power cord. If you'd like to be sure, take the
cover off the control box and measure voltage coming in from power
cord.
- If no reading (but know the outlet is good) the power
cord is bad.
- If you get a reading, it would seem the machine is a 'dead machine';
contact ChemFree tech support at 1-800-521-7182.